青光眼

OCT测量视网膜神经纤维层厚度及视盘参数在青光眼早期诊断中的应用

作者:韩瑞娟 等来源:首席医学网 发布日期:2010-02-08浏览:1726次

作者:韩瑞娟,具尔提·哈第尔    作者单位:(830001)中国新疆维吾尔自治区乌鲁木齐市,新疆医科大学第一附属医院眼科

【摘要】  青光眼是一种进行性视神经疾病,它能引起视神经结构改变,最终导致不可逆的视功能损害。青光眼的早期诊断对保护视功能有重要意义。光学相干断层扫描仪是一种准确性和可重复性高的视网膜神经纤维层厚度和视盘参数检测技术,为早期诊断青光眼,监测视神经损害及指导青光眼的治疗提供了新的思路。我们主要阐述该技术在青光眼早期诊断中的应用。

【关键词】  光学相干断层扫描;青光眼;早期诊断;视网膜神经纤维层厚度;视盘参数

OCT measurement of retinal nerve fiber layer thickness and optic disc parameters in early diagnosis of glaucoma

  RuiJuan Han, Juret Kadir

  Department of Ophthalmology, the First Affiliated Hospital of Xinjiang Medical University, Urumchi 830001, Xinjiang Uygur Autonomous Region, China

  Abstract

  Glaucoma is a progressive optic neuropathy characte rized by structural changes of the optic nerve, which may eventually develop into irreversible visual functional defects. Therefore, it needs early detection. Optical coherence tomography(OCT) is an accurate and reproducible method to measure retinal nerve fiber layer thickness and the optic disc parameters,it may show promise in early detection of glaucoma. This review primarily describes the technology in the early diagnosis of glaucoma.

  KEYWORDS:optical coherence tomography; glaucoma;early diagnosis;retinal nerve fiber layer thickness; optic disc parameters

  0引言
   
  青光眼是一种常见的致盲眼病,是一类以视网膜神经节细胞及其轴突的丧失为特征的视神经病变[1],其防治的关键在于早期检测、早期诊断及有效治疗。目前视野检查仍是青光眼诊断的金标准,但在自动视野计检测到最小暗点之前,通常有30%~50%的视网膜神经节细胞已经丢失[2],即视盘和视神经改变往往早于青光眼视野缺损[1],可表现为视盘形态和视网膜神经纤维层(retinal nerve fiber layer,RNFL)厚度的改变。虽然观察眼底视盘及其周围神经纤维层的形态学改变是青光眼早期诊断的重要方法之一,但该方法容易受到检查者主观因素的影响,检查结果变异较大。采用与计算机图像分析相结合的成像技术提供了客观、定量测量视盘参数和视盘周围RNFL厚度的可能性,也为早期诊断青光眼、监测视神经损害及指导青光眼的治疗提供了新的思路。光学相干断层扫描仪(optical coherence tomography,OCT)是一种利用光学相干原理设计的高分辨率的横截面断层扫描影像学检查方法,可客观、定量测量RNFL厚度及视盘参数,而且测量的准确度高[3]、可重复性好[4],能实时地显示视网膜的横断面图像,定量地测量各结构参数和变化,为疾病诊断提供客观依据。OCT在青光眼RNFL检测方面的应用已有诸多文献报道[5] ,且一致认为它能够区分健康人和青光眼患者,可早期敏锐地发现RNFL变薄或缺损及视盘盘沿组织进行性不对称丧失等病理改变[6]。目前临床上常用的OCT成像仪是由德国CarlZeiss公司生产的,它由低相干光干涉仪、监视器、眼底摄像机、计算机、显示器及彩色打印机组成[7],其轴向分辨率达10μm,横向分辨率达20μm,为OCT在青光眼诊断方面提供了更广泛的应用前景。我们主要探讨OCT测量RNFL厚度及视盘参数在青光眼早期诊断中的应用。

  1 RNFL厚度及视盘参数测量及扫描方法
   
  OCT检查在暗室中进行,检查前均用5g/L托吡卡胺滴眼液扩瞳(瞳孔直径>5mm)[8],检查者选择RNFL厚度扫描程序,OCT会自动给出注视光的位置,当被检眼视力较好时,使其注视注视光,称为内注视;当被检眼视力较差,使用外注视。一般来说,内注视比外注视重复性要好。常规和快速的RNFL厚度扫描是以视盘为中心,默认直径为3.4mm的连续环形扫描。分析扫描结果提供各眼视网膜反射图和RNFL厚度地形图、双眼TSNIT(颞、上、鼻、下4个象限RNFL厚度)曲线图及对比图。打印结果参数有12∶00方位的RNFL厚度(右眼为顺时针方向1∶00~12∶00各钟点位的RNFL厚度,左眼为右眼的镜像位),平均(Avg)、上方(S)、下方(I)、颞侧(T)及鼻侧(N)的RNFL厚度以及其他比值和参数。

  2 RNFL厚度及视盘参数测量的影响因素及可重复性

  2.1正常人RNFL厚度及视盘参数测量的变异和影响因素  OCT测量正常人的视盘周围RNFL厚度及视盘参数存在较大的变异,可能受到年龄、种族、屈光度数、眼轴长度、视盘面积大小等因素的影响[9]。(1)年龄和性别:Bowd等[10]研究发现RNFL厚度与年龄负相关(R2=0.17),与性别无明显相关。(2)种族:种族的不同可能造成RNFL厚度不同。在儿童和成人的研究中均发现白种人的RNFL厚度小于亚洲人或西班牙人[11,12]。(3)屈光不正和眼轴长度:虽然Bowd等[10]研究发现RNFL厚度与屈光度无明显相关,但大多数学者都认为随着正性屈光度数的减少和眼轴长度的增加, RNFL厚度有变薄趋势。这种趋势不仅存在于成人,也存在于儿童中[11,13]。(4)视盘面积:Bowd等[10]发现RNFL厚度与视盘大小正相关。Savini等[14]也发现视盘面积越小,RNFL厚度越薄,而视盘面积越大,RNFL厚度越厚。但是由于OCT是通过以视盘为中心,一定直径的环形扫描,而不是沿视盘边缘进行的扫描,所以他并不能确定这种变化是由于较大的视盘具有较多的视神经纤维所致,还是由于扫描圆离较大视盘的距离较近的缘故造成的。

  2.2 RNFL厚度及视盘参数测量的可重复性  影响OCT测量RNFL厚度可重复性的因素:(1)仪器型号:Bourne等[15]认为使用相同的扫描模式,OCT2所测得的RNFL厚度比OCT3厚。即使都使用OCT3,Sehi等[16]认为在不同仪器间测量结果存在着差异,这种差异在平均、上方RNFL厚度明显。(2)操作者:Villain等[17]认为操作者内的测量重复性较好,不同操作者间所测RNFL厚度在各象限及全周差异均无显著性。GürsesOzden等[18]认为不同操作者使用OCT3有好的重复性。(3)扫描方式和信号强度:Zafer等[19]报道,提高OCT3的轴向扫描组数可提高RNFL厚度测量的可重复性,且发现OCT3的轴向扫描组数介于100~256时所测RNFL厚度的可重复性最好。Budenz等[20]报道,用OCT3对正常人及青光眼患者行常规和快速RNFL厚度两种扫描程序进行重复性测量,两种扫描程序有类似的结果。GürsesOzden等[18]认为用OCT3测量正常眼时各扫描模式的重复性好。Wu等[21]认为用OCT3测量时,信号强度、低的分析可信度和较薄的RNFL厚度均可影响其可重复性。(4)瞳孔大小:尽管理论上OCT3可以在不扩瞳的基础上进行检查,Paunescu等[22]认为在扩瞳的条件下测量结果的重复性更好。Savini等[23]也发现瞳孔大小能轻度影响测量结果。但Zafer等[24]认为散瞳与否对检测结果无明显影响。(5)屈光间质:认为晶状体混浊会使RNFL厚度测量值下降。影响OCT检测视盘参数的可重复性的因素:(1)瞳孔大小:Paunescu等[22]用OCT3对10位正常志愿者于散瞳前后的视盘参数各进行3次重复性测量,结果显示扫描6次的所有视盘参数的ICC(intraclass correlation coefficient,等级内相关系数)值均>0.5,其中以杯盘比(C/D)为最好,ICC值为0.972,且比较散瞳前后扫描3次的所有视盘参数的ICC值差异无显著性。(2)参考平面:Leung等[25]报道用OCT设定在150,95,205μm 3种不同参考平面下检测视盘参数,以设定参考平面为150μm时所测视盘参数结果可重复性和准确性最好。

  3 OCT在青光眼早期诊断中的应用

  3.1 RNFL厚度及视盘参数测量与视野的关系  多项研究证实,在青光眼中RNFL厚度与视野损害密切相关,使用OCT检测到RNFL越薄,盘沿越窄,视野损害越明显,且视野损害与OCT图像中的RNFL变薄或缺损呈直接对应关系。Sihota等[26]发现在青光眼中所有RNFL厚度参数均与标准视野的平均缺损和矫正模式标准差明显相关。El Beltagi等[27]发现在青光眼患者中局部RNFL变薄与局部敏感度下降呈位置相关:OCT上6∶00~8∶00(下方和颞下方)的RNFL厚度改变和视野上方弓形和上方鼻侧阶梯区域的敏感度改变相对应。成云翠等[5]报道用OCT3测量平均RNFL厚度与视野检测的MD值的相关系数r=0.670。丁琦等[28]报道,原发性开角型青光眼患者MD与RNFL厚度呈负的直线相关(r=0.615,P=0.000)。Kanamori等[29]报道,在研究样本量较大的情况下,所得的平均RNFL厚度与视野的相关系数r=0.729。Leung等[25]有类似报道,他们通过对正常眼,开角型青光眼及可疑青光眼进行OCT和静态视野检测,所得的相关系数r值最大为0.75。OCT检测到的视盘参数与视野的相关性亦有报道:Schuman等[30]对一组青光眼患者用HRT和OCT3检测视盘,同时行SITA和FDT两种视野检查,然后将所测视盘参数与两种视野的MD值做相关分析,结果示SITA视野检查的MD值与HRT和OCT3相关性最好的参数,而与FPT视野检测的MD值相关性最好的都为CA。Leung等[31]对正常者,早期及中期青光眼患者行视盘和Humphrey242视野检测,将150,95,205μm 3种不同参考平面所得的视盘参数分别与MD做直线回归分析,结果发现3种参考平面下的所有视盘参数与MD都显著性相关,其中以RA(盘沿面积)相关性最好。成云翠等[32]报道OCT3能够检测到青光眼的早期视盘结构的改变,且与MD有基本一致的较好相关性。这些结果表明OCT3检测视盘参数与视野检查亦有很好的相关性。

  3.2 RNFL厚度及视盘参数测量在青光眼早期的应用  早期诊断是青光眼治疗的关键,有研究表明青光眼视野损害前6a即有视网膜神经纤维层的丢失,通过对RNFL厚度和视盘参数的测量能更早地发现青光眼。Leung等[31]报道,用OCT检测正常人、早期及中期青光眼患者视盘参数,结果示3组间所有视盘参数差异均有显著性(P<0.01)。季宝玲等[33]报道从正常人和青光眼、正常人和早期青光眼的组间均数比较和ROC曲线下面积两个角度分析,结果显示测量RNFL厚度有较强的青光眼早期诊断价值。Wang等[34]的研究发现,RNFL厚度的ROC曲线下面积较大(0.959~0.973),而与MD的相关系数r相对较小(0.158~0.557),这表示早期青光眼RNFL结构损害往往早于可探测到的视功能损害。Kim等[35]报道在正常眼压性青光眼患者中,未出现视野缺损的对侧眼中已有RNFL改变。Essock等[36]将早期青光眼患者和正常人RNFL厚度TSNI双峰曲线进行Fourier分析建立线性判别方程,结果示设定特异性为90%时,敏感性为76%,受试者操作特征曲线下面积为0.925。Choi等[37]报道高眼压症组下象限的平均RNFL厚度和正常组有较少的统计学差异,余象限差异无显著性,而青光眼组所有象限及全周的平均RNFL厚度值与高眼压症组和正常组间比差异有显著性(P<0.01)。视盘参数中的CD(视杯直径),CA(视杯面积),RA(盘沿面积),C/D AR(杯盘面积比)在高眼压症组和正常组间差异有显著性(P=0.045),而青光眼组与高眼压症组和正常组间差异有显著性(P<0.01)。
   
  总之,OCT因拥有分辨率高、非侵入性、可重复性强等优点,能真实反映青光眼RNFL厚度及视盘参数的改变,且具有较高的敏感性和特异性,为临床医生早期诊断青光眼,了解青光眼的严重程度提供帮助。虽然OCT技术近年来改进完善了不少,但OCT技术在青光眼早期诊断、病情跟踪和疗效评估中的应用还有待进一步研究。

【参考文献】
   1曹东,贺翔鸽,刘莛,等. OCT和SLP定量检测视网膜神经纤维层的研究进展.国际眼科杂志2008;8(3):571574

  2季宝玲.光学相干断层成像术测量视网膜神经纤维层厚度在青光眼早期诊断中的意义.国际眼科杂志2007;7(4):10191021

  3黄丽娜, Schuman J, Wang N.光学相干断层扫描与组织形态学检测猴青光眼视网膜神经纤维层厚度的比较.中华眼科杂志2001; 37(3):188193

  4刘杏,凌运兰,周文炳,等.光学相干断层成像术对原发性开角型青光眼视网膜神经纤维层的定性和定量检测.中华眼科杂志2000; 36(6):420424

  5成云翠,段宣初.HRT2Ⅱ和OCT3检测视网膜神经纤维层厚度在原发性开角型青光眼早期诊断中的应用.国际眼科杂志2007;7(4):10221024

  6成云翠,段宣初,高莉茉,等.OCT3和HRTⅡ检测视盘参数在青光眼早期诊断中的应用价值.眼视光学杂志2007;9(6):364368

  7成云翠,胡妙妙,段宣初.Stratus OCT在青光眼早期诊断和随访中的应用.国际眼科杂志2006;6(3):671675

  8周晓晴,钟一声. OCT在正常眼压性青光眼患者中的应用研究.国际眼科杂志2008;8(8):15871590

  9曹斯嘉,张士胜,王玲.OCT测量视网膜神经纤维层厚度在青光眼诊断中的研究进展.国际眼科杂志2008;8(8):16341636

  10 Bowd C, Zangwill LM, Blumenthal EZ, et al. Imaging of the optic disc and retinal nerve fiber layer:the effects of age, optic disc area, refractive error and egender. J Opt Soc Am A 2002;19(1):197 207

  11 Hougaard JL, Ostenfeld C, Heijl A, et al. Modelling the normal retinalnerve fibre layer thickness as measured by Stratus optical coherence tomography. Graefes Arch Clin Exp Ophthalmol 2006;244 (12): 16071614

  12 Salchow DJ, Oleynikov YS, Chiang MF, et al. Retinal nerve fiber layer thickness in normal children measured with optical coherence tomography. Ophthalmology 2006;113 (5):786791

  13 Huynh SC, Wang XY, Rochtchina E, et al. Peripapillary retinal nerve fiber layer thickness in a population of 6yearold children:findings by optical coherence tomography. Ophthalmology 2006; 113 (9):15831592

  14 Savini G, Barboni P, Carbonelli M, et al. The effect of scan diameter on retinal nerve fiber layer thickness measurement using stratus OCT. Arch Ophthalmol 2007;125 (7):901905

  15 Bourne RR, Medeiros FA, Bowd C, et al. Comparability of retinal nerve fiber layer thickness measurements of optical coherence tomography instruments. Invest Ophthalmol Vis Sci 2005;46 (4):12801285

  16 Sehi M, Guaqueta DC, Feuer WJ, et al. A comparison of structural measurements using Stratus optical coherence tomography instruments. J Glaucoma 2007;16(3):287292

  17 Villain MA, Green field DS. Peripapillary nerve fiber layer thickness measurement reproducibility using optical coherence tomography. Ophthalmic Surg Laser Imaging 2003;34(1):3337

  18 CursesOzden R, Teng C, Vessani R, et al. Macular and retinal nerve fiber layer thickness measurement reproducibility using optical coherence tomography (OCT3). J Glaucoma 2004;13(3):238244

  19 Zafer S, CursesOzden R, Makornwattana M, et al. Scanning protocal choice effects optical coherence tomograpgh (OCT3) measurements. J Glaucoma 2004;13(2):142144

  20 Budenz DL, Chang RT, Huang X, et al. Reproducibility of retinal nerve fiber thickness measurements using the Stratus OCT in normal and glaucomatous eyes. Invest Ophthalmol Vis Sci 2005;46(7):24402443

  21 Wu Z, Vazeen M, Varma R, et al. Factors associated with variability in retinal nerve fiber layer thickness measurements obtained by optical coherence tomography. Ophthalmology 2007;114(8):15051512

  22 Paunescu LA, Schuman JS, Price LL, et al. Reproducibility of nerve fiber thickness,macular thickness, and optic nerve head measurements using stratus OCT. Invest Ophthalmol Vis Sci 2004; 45 (6):17161724

  23 Savini G, Zanini M, Barboni P, et al. Influence of pupil size and cataract on retinal nerve fiber layer thickness measurements by Stratus OCT. J Glaucoma 2006;15(4):336340

  24 Zafar S, CursesOzden R, Vessani R, et al. Effect of pupillary dilation on retinal nerve fiber layer thickness measurements using optical coherence tomograpgh. J Glaucoma 2004;13(1):3437

  25 Leung CK, Yung WH, Ng AC, et al. Evaluation of scanning resolution on retinal nerve fiber layer measurement using optical coherence tomograpgh in normal and glaucomatous eyes. J Glaucoma 2004;13(6):479485

  26 Sihota R, Sony P, Gupta V, et al. Diagnostic capability of optical co herence tomography in evaluating the degree of glaucomatous retinal nerve fiber damage. Inves Ophthalmol Vis Sci 2006;47 (5):20062010

  27 El Beltagi TA, Bowd C, Boden C, et al. Retinal nerve fiber layer thickness measured with OCT is related to visual function in glaucomatous eyes. Ophthalmology 2003;110 (11):21852191

  28丁琦,陆豪,严良,等.OCT检测原发性开角型青光眼视网膜神经纤维层厚度与视野平均缺损的相关性分析.临床眼科杂志2006;14(4):355356

  29 Kanamori A, Nakamura M, Escano MF, et al. Evaluation of the glaucomatous damage on retinal nerve fiber layer thickness measurement by optical coherence tomography. Am J Ophthalmol 2003;135(4):513520

  30 Schuman JS, Wollstein G, Farra T, et al. Comparison of optic nerve head measurements obtained by optical coherence tomography and confocal scanning laser ophthalmoscopy. Am J Ophthalmol 2003;135(4): 504512

  31 Leung CK, Chan WM, Hui YL, et al. Analysis of retinal nerve fiber layer and optic nerve head in glaucoma with different reference plane offsets,using optical coherence tomography. Invest Ophthalmol Vis Sci 2005;46(3):891899

  32成云翠,段宣初. OCT3检测视盘参数与视野的相关性研究.国际眼科杂志2008;8(7) : 16121614

  33季宝玲,游逸安,方爱武,等.OCT测量视网膜神经纤维层和黄斑厚度在青光眼早期诊断中的意义. 眼视光学杂志2008;10(1):5458

  34 Wang JJ, Liu X, Zen YF, et al. Comparison of two different scanning protocols of retinal nerve fiber blayer thickness. Ophthalmic CHN 2005;14(6):400404

  35 Kim DM, Hwang US, Park KH, et al. Retinal nerve fiber layer thickness in the fellow eyes of normaltension glaucoma patients with unilateral visual field defect. Am J Ophthalmol 2005;140 (1):165166

  36 Essock EA, Sinai MJ, Bowd C, et al. Fourier analysis of optical coherence tomography and scanning laser polarimetry retinal nerve fiber layer measurements in the diagnosis of glaucoma. Arch Ophthalmol 2003;121(9):12381245

  37 Choi MG, Han M, Kim YI, et al. Comparison of glaucomatous parameters in normal ocular hypertensive and glaucomatous eyes using optical coherence tomography 3000. Korean J Ophthalmol 2005;19(1):4046

>>只有眼网认证的眼科医生用户,才能阅读全部内容,请登录:
用户名:  密码:  验证码:  
>>如果您还不是眼网认证眼科医生,欢迎您免费注册:
中国眼网郑重声明: 注明“来源:中国眼网”的作品,版权均属于中国眼网所有,未经中国眼网授权不得转载、摘编或以其它方式使用上述作品。已经中国眼网授权使用的作品,应在授权范围内使用,并注明“来源:中国眼网”。违反上述 声明者,中国眼网将追究其相关法律责任。
凡中国眼网,注明“来源:XXX(非中国眼网)”的作品,均转载自其它媒体,转载目的在于传递更多信息,并不代表中国眼网赞同其观点,同时中国眼网对其真实性不负责。
如因作品内容、版权及其它问题需同中国眼网联系的,请在30日内发送EMAIL至:eyenet@yahoo.cn

评论数据更多评论

现在有 人对本文发表评论
正在获得评论数据,请稍后...

发表评论

姓名: E-mail: 验证码: 您还可以输入500个字符
如果要咨询眼病相关问题,请点击按钮 
  • 注意:您发表的评论需要经过我们的审核后,才能显示
Copyright © 2006-2008 www.eyenet.com.cn 版权所有 中国眼网  中国眼科第一门户  辽ICP备05019381号 未经授权请勿转载
地址:中国沈阳 中国眼网业务联系电话:024-88062461 中国眼网业务信箱: 业务QQ:暂未开通(MON-FRI 8:00-17:00)